Mode Type: 
Whole Blood, Predilute, Capillary
Autoloader: 
60 Tubes; Continuous Loading with Auto-Mixing 
36 Parameters: 
CBC: WBC, RBC, Hb, HCT, MCV MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW-SD, PDW-CV, P-LCR, P-LCC, DIFF: Neutrophils % & #, Lymphocytes % & #, Monocytes % & #, Eosinophils % & #, Basophils % & # *ALY % & #, *LIC % & #, *NLR, NRBC% & #, CWBC, *Mentzer index and *RDWI *Research use only
Principle of Measurement: 
RBC/PLT/WBC : Electrical Impedance
WBC DIFF: Tri-angle Laser Flowcytometry
Hb: Cyanide Free Colorimetry
MCV: Measured
HCT: Calculated
Histograms :
3 (RBC/WBC/PLT)
Scattergrams :
3 (2D Scattergram), 1 (3D Scattergram)
QR Code Reader:
For reading sample bar codes & direct input of QC values
Sample Volume: 
Whole Blood: 20 µL
Pre- diluted: 20 µL
Capillary: 20 µL
CBC Mode : 13 µL
Linearity Range:
WBC : 0 - 300 ( x 103/µL)  
RBC : 0 - 300 ( x 103/µL)  
Hb: 0 - 25.0 (g/dL)
HCT: 0 - 67 (%)
PLT : 0 - 3000 (x 103/µL)
Calibrator:
ELite H Cal 
Tri-level Controls:
Erba H5 CON L, N, H
Open Vial Stability at 2 - 8˚C: 16 Days
Throughput: 
80 Tests/Hr 
Quality Control:
500 files with L-J Graph
X- Bar Analysis
Data Storage: 
1,00,000 Results with Graphs 
Dimension (mm): 
650x550x610 
Weight (kgs): 
59
Reagents:
ELite 580 Dil
ELite H580Lyse1
ELite H580Lyse2
ELite H580Lyse3
Elite H Clean
Interfaces: 
4 USB ports ( for connecting external devices including external printers) + 1 LAN Port (HL 7, TCP/IP; LIS connectivity)
Operating Environment Temperature: 
15-40˚C
Humidity:
 30-90%
Power Requirement:
A.C. 100 - 240 V with built-in voltage stabiliser 