Hb-Vario :: Transasia Bio-Medicals

Hb-Vario

Fully automated HbA1c testing systems based on HPLC technology.

Product Description:

Hb-Vario is a true HPLC platform for HbA1c testing with HbA2/F flagging. The outstanding feature of the Hb - Vario is, its unique CLE technology, which enables complete elimination of labile glycohemoglobin fraction to ensure accurate quantitation of HbA1c results. The results are available in 3.5 min.

Product Features:

  • Better precision, CV < 2.5 %
  • Patented CLE technology enables complete elimination of labile glycohemoglobin ensuring accurate HbA1c results.
  • HbA1c test results in 3.5 minutes with HbA2/HbF flagging.
  • Separation of hemoglobin fractions
  • Complete removal of labile fraction up to 10%
  • No interference of HbF up to 10%
  • No interference of lipemia or icterus, can run on non-fasting patient samples
  • HbA1c reportability presence of HbS, HbC, HbD and HbE
  • Compact bench top model
  • User friendly touch screen operation
  • Sampling from multiple sized primary tubes, including microcapillary collection tubes
  • On board QC functions with Levey-Jennings and statistical calculations
  • Unlimited storage of calibration, QC data and chromatograms
  • No sample preparation required
  • Simple reagent and column replacement
  • Automated maintenance
  • RFID reagent identification and tracking
  • Automatic barcode identification of samples
  • NGSP certified
  • Calibrators and Controls IFCC traceable
  • Complete kit with buffers, calibrator and column

Sampler Handler: 10 Position Rack Loader, positive identification by internal barcode reader

Sample Type: Whole blood on EDTA K3 or K2 anticoagulant Haemolysate from EDTA whole blood samples

Time for Analysis: Throughput approximately 3.5 minutes between samples

Interferences: Complete removal (not separation) of labile up to 10%

Sampling System: Sample Aspiration Probe coupled with a high precision,

Sample Aspiration Volume: 14.5 µl in whole blood, 670 µl in lysate mode.

Pump: 2 high pressure syringe pumps

Detector Type: UV-Vis

Reagent Traceability: RFID

Printer: Built- in thermal printer with possibility to connect to external printer

Reagents used with this 'Hb-Vario'