XL 640 CRS :: Transasia Bio-Medicals

XL 640 CRS

XL 640 CRS is a Fully Automated Clinical Chemistry Analyzer with Autoloader

Product Description:

XL-640 with CRS is a random access, discrete, patient prioritized, fully automated clinical chemistry analyzer with autoloader designed to meet the needs of today’s modern clinical laboratories with high workload. Once samples are loaded, it is a complete walk-away system. XL 640 CRS is designed to operate smoothly over the long haul and to deliver consistent quality report.

Product Features:

  • Throughput of 400 tests/ hour (Photometric) and 640 tests/hour with ISE
  • 56 Reagent Positions with On-board cooling for reagents
  • 162 Sample positions with /Autoloader
  • Separate probes for reagent R1 & R2 and dedicated probe for sample
  • Clot detection for sample probe, Reflex testing and Serum Indices identification
  • 72 permanent hard glass cuvettes with On-board Laundry
  • Automatic dilution for high abnormal samples & Auto Rerun facility
  • Extensive QC Programmes with Levy –Jennings Curve & QC rules. Automatic calculation for Mean SD & CV%
  • Direct loading for primary tubes
  • Touch screen User Friendly GUI with Bi-directional connectivity.

Product Applications

  • Routine Chemistries: LFT, KFT, Lipids, Diabetic Profile, Ions, Anaemia profile, many more parameters available along with controls and calibrators
  • Special Chemistries: HbA1c, Iron, UIBC, Ferritin, RF, CRP, ASO, MAL, Glucose with Hexokinase Method.

System General

System Type: Automated, Discrete, Random Access, Patient Prioritized Clinical Chemistry Analyzer with Autoloader

Throughput: 400 photometric tests/hour & 640 tests/hour with ISE (*optional ISE with Na, K, Cl, Li)

Analytical Methods: 1-Point, 2-Point, Rate-A, Rate-B, Direct Potentiometry(optional)

Barcode Reader: Both for Samples & Reagents (Optional)

Reactions Tray: 72 hard glass cuvettes

Reaction Mixing: Dual Stirrer with variable speed

Reading Volume: 180 µl

On-Board Laundry: 6 stage cleaning, 2 stage drying

Photometer: Multi-wavelength diffraction grating with 12 wavelengths (340, 376, 415, 450, 480, 505, 546, 570, 600, 660, 700 & 750 nm)

O.D. Range: 0 - 3.0

Light Source: Halogen Lamp

Detector: Silicon Photodiodes

Water Consumption: < 14 liters/hour

Programmable Parameters: Default system pack parameters + upto 99 user defined parameters, Unlimited profiles and unlimited calculation items

Quality Control: Q.C. plot data with Q.C. Rules. Provision for lab mean. Twin plot.

Caliberation: K- factor, Linear (1,2 point and multipoint) 4P and 5P Logit-log, Cubic spline,exponential, polynomial. On-board serial dilution for calibrator

Power Requirement:

AC 220 V ± 10 %, 50 Hz or

AC 110 V ± 10 %, 60 Hz

Power Consumption: 800 VA

Dimensions: Approximately 897 mm (w) x 655 mm (D) x 1170 mm (H)

Weight: Approximately 200 Kg

Sample Handling

Sample: Serum, Plasma, Urine, CSF, Whole blood, other

Sample Unit: upto 162 positions for samples/STAT/calibrator/controls with Autoloader

Sample Pipetting: 2-70 µl (adjustable in 0.1 µl steps) for biochemistry,70 µl fixed for ISE capacitance probe with liquid level sensing and vertical obstruction detection , clot detection and serum indices and Reflex Testing

Auto Rerun: Repeat with same, increade and decresaed volume (upto 1:150)

Sample Tubes/Cups: Primary tubes of 5 ml, 7ml and 10 ml and sample cups

Reagent Handling

Reagent Tray: 56 positions for reagents with on-board cooling

Reagent Pipetting:

R1: 60 - 300 µl (Adjustable in 1 µl step)

R2: 0 or 10 - 300 l (Adjustable in 1 µl step)

Capacitance probe with level sensing and vertical obstruction detection

System Interface

Analyzer-PC: USB

PC- HOST Computer: Bidirectional TCP/IP and RS - 232

PC- Printer: USB

Operating System: Windows based with Touch Screen User Friendly GUI

Database: Unlimited Results

Reagents used with this 'XL 640 CRS'