Hardware Specifications:
System type: Open, Flowcell
Measurement Principle: Colorimetry (Rate/End Point)
Applicable Analytes: Photometric Assays - Substrates, Enzymes, Lipid, Protein, Sugar, Inorganic Substances
Setup: Benchtop
Lightsource: Quartz Halogen Lamp, 12V, 20 Watts
Photometric Range: Range 340-670nm
Resolution: 0.0001
Optics: 8-wavelength static photometric group with the following interference filters: 340, 405, 450, 505, 546, 578, 600 and 670nm
DETECTORS: 6 UV/ VIS Silicon Photodiodes
CUVETTE: Unique Triple Cuvette System:
- 18 microlit. flow cell
- 10 ml rectangular polystyrene cuvette
- 6mm round Glass Tube
THERMOSTAT: Peltier Temperature Control Range 20 degrees C to 40 degrees C with 1 degree C steps
SAMPLE ASPIRATION: Using a peristaltic pump, Aspiration volume programmable 100 microlit. to 2997 microlit.
PRECISION: Better than 1% CV
KEYBOARD: Rugged waterproof, non-tactile legend membrane panel with 25 keys
SMPS: EXTERNAL (24W; 60W)
PRINTER: High resolution, 384 dots per line, thermal type with full graphics facility and option for connecting an external USB printer
DISPLAY: High-resolution Graphics LCD 240x64 dots with backlit, View area 134mm x 40mm
INTERFACES: USB B type port for the host computer and USB A type port for External USB Jet Desk Printer. External PT/ AT keyboard
PROCESSOR: 32-bit Cortex M4 Microcontroller
MEMORY: Built-in 1MB Flash, 136KB SRAM, 16KB EEPROM, and 64 KB ROM
REAL-TIME CLOCK: Built-in
REPLACEABLE BATTERY ON PCB: Lithium battery CR2032 (3V)
WASTE COLLECTION: Waste collected in closed container
INCUBATOR (OPTIONAL): 16 incubating positions
OPERATING TEMPERATURE: 20 degrees to 40 degrees C
STORAGE TEMPERATURE: 5 to 45 degrees C
HUMIDITY: Max 80% RH, non-condensing
MAINS SUPPLY: 115/230 V AC; 50/60 Hz. (input for power adapter)
SIZE INSTRUMENT: 446 (W) x 312 (D) x 175 (H) mm
WEIGHT: Approximately 6kg.
*NOTE: Above specifications subject to change without prior notice.
Software Specifications:
11 analytical modes of operation:
- 1-point Linear
- 2-point Linear
- Rate A Linear
- 1-point Non-Linear (Cubic Spline/ Point to Point)
- 2-point Non-Linear (Cubic Spline/ Point to Point)
- Rate-A Non-Linear (Cubic Spline/ Point to Point)
- 1-point Sample Blank Linear
- 1-point Sample Blank Linear (Cubic Spline/ Point to Point)
- Absorbance
- Index Mode
260 totally "OPEN" test programs selectable through the keypad
Parameters can be viewed, edited, and printed
Facility to memorize Reagent Blank ABS, Sample Blank ABS, Standard ABS, Factor, Non-linear curve
Patient reports for the last 10,000 results stored in memory are obtained by
- Date
- ID
- Both by Date and ID
Graphic presentation of Non-Linear calibration curves
QC (QC data available for three controls per test)
Rate calculation by the least square regression method
Automatic zeroing of Water OD during Run Test
Total reaction time: Delay time and read time both are programmable between 1 second to 999 seconds
Non-Linear Calibration (6 standards)
On-line reagent stability check
Substrate depletion/ non-linearity check
In-built reaction non-linearity check
Monochromatic as well as bichromatic measurements are possible
Calibration of peristaltic pump
Online display of cuvette temperature
Self-diagnostics
Help messages in erroneous entry
Reports can be printed on a USB printer or thermal printer by either using the Print key or automatic
USB-based barcode scanner for sample entry and search
Incubator (Optional)