ECL 760 ® :: Transasia Bio-Medicals

ECL 760 ®

Seven Channel Fully Automated Random Access Analyzer for Complete Hemostasis

Product Description:

It’s an advanced, fully automated random access analyzer for Complete Hemostasis. This seven channel coagulation analyzer redefines the benefits in its class. This system is ideal for mid-high range coagulation workload, especially for routine PT, aPTT and some rare parameters.

Product Features:

  • Interactive touch screen interface/ keyboard
  • Clotting Graph provides details of the clotting test
  • Cuvette and reagent management software
  • Twelve levels of QC with Levy-Jennings Plot- Identifies random and systematic errors and multirules.
  • Temperature stability at 37 degree C to perform the tests
  • Low Reagent Consumption
  • Windows software
  • CE mark

Product Applications:

  • Clotting Tests such as PT, aPTT, Fibrinogen, Thrombin time, Factors, Lupus anticoagulant, Protein S can be performed using Light scatter principle
  • Protein C, Antithrombin III can be done by Chromogenic Principle
  • Tests like D Dimer can be performed using Immuno-turbidimetry principle

General Specifications:

  • 7 channel fully automated hemostasis instrument
  • It can perform 10 tests per profile
  • Sample id can be 14 digits numeric, self generating id or custom id
  • It is also possible to have continuous loading of samples 
  • It can run assays in single or duplicate (programmable per assay)
  • The System uses the much proven light scatter principle of detection

Tests principles:

  • Clotting (scattered light at 640nm)
  • Chromogenic (colorimetry at 405nm)
  • Immuno-turbidimetric (turbidimetry at 575nm)

Dimensions / Weight:

  • Dimension :660 x 580 x 510 mm
  • Weight : ~53 kg

Environmental requirements:

  • Operating Temperature : 17-32°C
  • Humidity: Max. 80% Relative Humidity, non-condensing

Power:

  • Power Supply : 100-240 VAC and 50/60 Hz
  • Power Consumption : 45 Watts
  • In-Rush Power : 150 VA or less

Calibrations:

  • Up to 6 calibration levels per assay
  • Storage of the active calibration curve

Quality Controls:

  • Up to 12 different control total can be assigned to any given assay.
  • Automatic Levey-Jennings plotting
  • Multi Rules
  • Storage of Quality Control results

Stored Results:

  • Independent storage allocation in the instrument memory for Quality Control (QC) and patient determinations
  • Up to100 results for QC determinations (run in duplicate)
  • Up to 100000 results for patients (run in duplicate)
  • New results automatically replace the oldest results of the appropriate category past the maximum storage allowance
  • Reaction curves are stored

Reagents used with this 'ECL 760 ®'